GW Instek GPT-9601
100VA, AC/DC/IR Electrical Safety Tester
Search
Manufacturers
100VA, AC/DC/IR Electrical Safety Tester
Selection Guide of the GPT-9600 Series | |||||
Model | GPT-9601 | GPT-9602 | GPT-9612 | GPT-9603 | |
Output Capacity | 100VA | ||||
AC Withstanding (ACW) | 5kVac 20mA | 5kVac 20mA | 5kVac 20mA | 5kVac 20mA | |
DC Withstanding (DCW) | ------ | 6kVdc 6mA | ------ | 6kVdc 6mA | |
Insulation Resistance (IR) | ------ | ------ | 50 /100/250/ 500/1000Vdc 2000MΩ | 50/100/250/ 500/1000Vdc 2000MΩ | |
Ground Bond (GB) | ------ | ------ | ------ | ------ | |
ARC Detection | YES | YES | YES | YES | |
Ramp-Up Function | 0.1s Fixed | 0.1s Fixed | 0.1s Fixed | 0.1s Fixed | |
Memory Storage for Setting | 1 | 2 | 2 | 3 | |
AUTO Mode | ------ | ------ | AC-IR or IR-AC | AC-IR or IR-AC DC-IR or IR-DC | |
Safety Interlock | YES | YES | YES | YES | |
Remote Terminal | YES | YES | YES | YES | |
Signal I/O | YES | YES | YES | YES | |
USB/RS232C Interface | ------ | ------ | ------ | ------ | |
GPIB Interface | ------ | ------ | ------ | ------ | |
Display, LCD | 240 x 48 | 240 x 48 | 240 x 48 | 240 x 48 | |
CE Compliance | YES | YES | YES | YES |
AC Withstanding Voltage [for all models] | |||
Output-Voltage Range | 0.10kV~ 5.00kV | ||
Output-Voltage Resolution | 10V | ||
Maximum Rated Load | 100 VA (5kV/20mA) | ||
Maximum Rated Current | 20mA (0.5kV | ||
Output-Voltage Waveform | Sine wave | ||
Output-Voltage Frequency | 50 Hz / 60 Hz selectable | ||
Voltage Regulation | ± (1.5% + 2 counts) [full load ->no load] | ||
Voltmeter Accuracy | ± (1.5% of reading + 2 counts) | ||
Current Measurement Range [1] | 0.01mA~20.0mA | ||
Current Best Resolution | 0.01mA(0.01mA~9.99mA) 0.1mA(10.0mA~20.0mA) | ||
Current Measurement Accuracy | ± (2.0% of rdg + 5 counts) when HI SET<1.00mA ± (2.0% of rdg + 3 counts) when HI SET=1.00mA | ||
Window Comparator Method | Yes | ||
ARC Detect | Yes | ||
RAMP (Ramp-Up Time) | 0.1s Fixed | ||
TIMER (Test Time) | OFF, 1s~180s | ||
GND | ON | ||
[ 1 ] Output Limitation in AC Withstanding Voltage Testing as below. | |||
Upper Current | Pause | Output Time | |
AC | 15mA?I?20mA | At least as long as the output time | Approx. 180 seconds |
0.01mA?I<15mA | Not necessary | Continuous output possible |
DC Withstanding Voltage [for GPT-9602 / 9603] | |||
Output-Voltage Range | 0.10kV~ 6.00kV | ||
Output-Voltage Resolution | 10V | ||
Maximum Rated Load | 25W (5kV/5mA) | ||
Maximum Rated Current | 6mA (0.5kV | ||
Voltage Regulation | ± (1.5% + 2 counts) [full load ->no load] | ||
Voltmeter Accuracy | ± (1.5% of reading + 2 counts) | ||
Current Measurement Range [1] | 0.01mA~6.00mA | ||
Current Best Resolution | 0.01mA(0.01mA~6.00mA) | ||
Current Measurement Accuracy | ± (2.0% of rdg + 5 counts) when HI SET<1.00mA ± (2.0% of rdg + 3 counts) when HI SET=1.00mA | ||
Window Comparator Method | Yes | ||
ARC Detect | Yes | ||
RAMP (Ramp-Up Time) | 0.1s Fixed | ||
TIMER (Test Time) | OFF, 1s~180s | ||
GND | ON | ||
[ 1 ] Output Limitation in DC Withstanding Voltage Testing as below. | |||
Upper Current | Pause | Output Time | |
DC | 0.01mA?I?6.00mA | Not necessary | Continuous output possible |
Insulation Resistance | [for GPT-9612 / 9603] | |
Output Voltage | 50V/100V/250V/500V/1000V | |
Output-Voltage Accuracy | ± (3.0% of setting + 1 count) with no load | |
Voltage Regulation | ± (1.5% + 2 counts) [full load ->no load] | |
Voltmeter Accuracy | ± (1.5% of reading + 2 counts) | |
Resistance Measurement Range | 1MO~ 2000MO | |
Test Voltage | Measure Range | Accuracy |
50V / 100V / 250V | 1~50MΩ 51~2000MΩ | ± (5% of reading + 2MΩ) ± (10% of reading + 2MΩ) |
500V / 1000V | 1~500MΩ 501~2000MΩ ± (5% of reading + 2MΩ) | ± (10% of reading + 2MΩ) |
Output Impedance | 600kΩ | |
Window Comparator Method | Yes | |
RAMP (Ramp-Up Time) | 0.1s Fixed | |
TIMER (Test Time) | 1s~180s | |
GND | OFF | |
General | ||
Note | The specifications apply when the GPT-9600 series are powered on for at least 30 minutes at 15°C~35°C. | |
Display | 240 x 48 Ice Blue dot matrix LCD | |
AUTO Mode* | AC-IR / IR-AC / DC-IR / IR-DC | |
Interface | Remote, Signal I/O | |
Power Source | AC 100-120 V / 220-240 V ±10% | |
Power Line Frequency | 50/60Hz | |
Power Consumption | Max 400VA | |
Dimensions | 322(W) x 148(H) x 385(D) mm | |
Weight | Approx. 9kg max. | |
* The available Auto mode depends on the selected model | ||
Environment | ||
Note | To operate over the specific temperature range may cause damage to the circuit. And, do not use in a place where strong magnetic or electric field exists. | |
Warranty Environment | 15°C to 35°C, Humidity 70% (no condensation) | |
Operation Environment | 0°C to 40°C, Humidity 70% (no condensation) | |
Storage Environment | -10°C to 70°C, Humidity 85% (no condensation) | |
Installation Location | Indoors at an amplitude of up to 2000m |