Search

Categories

  New 
  Used 

Manufacturers

GW Instek GDM-8341

Stationary digital multimeter GDM-8341

More details

Features
※ 50,000 counts, VFD display
※ Dual measurement/Dual display
※ Selectable measurement speeds, the maximum: 40 Readings/s for DCV
※ The basic precision of DC voltage: 0.02% 
※ Auto/manual Ranging 
※ True RMS (AC,AC+DC) measurements
※ 11 different measurement functions
※ Max./Min., REL, MX+B, 1/X, Ref%, Compare, Hold, dB, dBm
※ Standard USB device interface to connect with a computer
※ Temperature measurement function (GDM-8342 only)
※ USB storage for Data collection (GDM-8342 only)
※ Optional GPIB (factory install for GDM-8342)

 

Note: The specifications apply when the DMM is warmed up for at least 30 minutes and operates only under main display in slow mode.

 

FunctionRange (*2)ResolutionInput Resistance
etc.
Accuracy(*1)
1 Year 23°C±5°C
DC Voltage500.00 mV10μV10MΩ or >10GΩ0.02 + 4
5.0000 V100μV10MΩ or >10GΩ0.02 + 4
50.000 V1mV11.11MΩ0.02 + 4
500.00 V10mV10.1MΩ0.02 + 4
1000.0 V100mV10MΩ0.02 + 4
Resistance500.00 Ω10mΩ0.83mA0.10 + 5 (*3)
5.0000 kΩ100mΩ0.83mA0.10 + 3 (*3)
50.000 kΩ83μA0.10 + 3
500.00 kΩ10Ω8.3μA0.10 + 3
5.0000 MΩ100Ω830nA0.10 + 3
50.000 MΩ1kΩ560nA//10 MΩ0.30 + 3
DC Current500.00 μA10nA0.06V max.0.05 + 5
5.0000 mA100nA0.6V max.0.05 + 4
50.000 mA1μA0.14V max.0.05 + 4
500.00 m A10μA1.4V max.0.10 + 4
5.0000 A100μA0.5V max.0.25 + 5
10.000 A1mA0.8V max.0.25 + 5
Continuity5000.0 Ω100mΩ0.83mA0.10 + 5
Diode Test5.0000 V100μV0.83mA0.05 + 5
Capacitance5.000 nF: 0.5~1nF0.001nF8.3μA2.00 + 20
5.000 nF: 1~5nF0.001nF8.3μA2.00 + 10
50.00 nF: 5~10nF0.01nF8.3μA2.00 + 30
50.00 nF: 10~50nF0.01nF8.3μA2.00 + 10
500.0 nF0.1nF83μA2.00 + 4
5.000 μF1nF0.56mA2.00 + 4
50.00 μF10nF0.83mA2.00 + 4
True RMS Volatge 
AC (or AC+DC) (*4/*5/*6)
500.00 mV10μV30Hz – 50Hz1.00 + 40
50Hz – 10kHz0.50 + 40
10kHz – 30kHz2.00 + 60
30kHz – 100kHz3.00 + 120
5.0000 V100μV30Hz – 50Hz1.00 + 20
50Hz – 10kHz0.35 + 15
10kHz – 30kHz1.00 + 20
30kHz – 100kHz3.00 + 50
50.000 V1mV30Hz – 50Hz1.00 + 20
50Hz – 10kHz0.35 + 15
10kHz – 30kHz1.00 + 20
30kHz – 100kHz3.00 + 50
500.00 V10mV30Hz – 50Hz------
50Hz – 10kHz0.50 + 15
10kHz – 30kHz1.00 + 20
30kHz – 100kHz3.00 + 50
750.0 V100mV30Hz – 50Hz------
50Hz – 10kHz0.50 + 15
10kHz – 30kHz------
30kHz – 100kHz------
True RMS Current
AC (or AC+DC) (*4/*5/*6)
500.00 μA10nA30Hz – 50Hz1.50 + 50
50Hz – 2kHz0.50 + 40
2kHz – 5kHz1.50 + 50
5kHz – 20kHz3.00 + 75
5.0000 mA100nA30Hz – 50Hz1.50 + 40
50Hz – 2kHz0.50 + 20
2kHz – 5kHz1.50 + 40
5kHz – 20kHz3.00 + 60
50.000 mA1μA30Hz – 50Hz1.50 + 40
50Hz – 2kHz0.50 + 20
2kHz – 5kHz1.50 + 40
5kHz – 20kHz3.00 + 60
500.00 mA10μA30Hz – 50Hz1.50 + 40
50Hz – 2kHz0.50 + 20
2kHz – 5kHz1.50 + 40
5kHz – 20kHz3.00 + 60
5.0000 A100μA30Hz – 50Hz2.00 + 40
50Hz – 2kHz0.50 + 30
10.000 A1mA30Hz – 50Hz2.00 + 40
50Hz – 2kHz0.50 + 30
Frequency10Hz – 500Hz------------0.01 + 5
500Hz – 500kHz0.01 + 3
500kHz – 1MHz0.01 + 5
Temperature (*7)
(Thermocouple)
-200 °C ~ +300 °C0.1 °CJ / T / K2 °C (typical)
DisplayVFD, Tow Colors Display
InterfaceUSB Device, USB Host (for GDM-8342 only), GPIB (optional for GDM-8342 only)
Power SourceAC 100 V / 120 V / 220 V / 240 V ±10%, 50-60Hz
Power ConsumptionMax. 15VA
Dimensions265(W) x 107(H) x 302(D) mm
WeightApprox. 2.9kg without option

 

1. Accuracy = +/- (% of reading + digits)
2. 2% overrange on all ranges, except 10A is 20% overrange
3. REL function is on
4. The AC+DC is in AC coupled
5. The accuracy of AC+DC is equal to AC with 10 more digits added
6. AC characteristics are for sinewave input > 5% of range
7. Specifications do not include probe accuracy

30 other products in the same category:

Previous

Specials

» All specials

Contact us

Our hotline is available from Monday to Friday

+48 58 742 11 27

info@LabAparaty.pl