GW Instek LCR-8105G
LCR meter LCR-8105G with graphic (20Hz ~ 5MHz), RS-232, GPIB, DC measure
Search
Manufacturers
LCR meter LCR-8105G with graphic (20Hz ~ 5MHz), RS-232, GPIB, DC measure
The LCR-8000G Series LCR meter, with test frequency up to 10MHz, provides accuracy, versatility and high resolution for a wide range of component measurements, even including DC resistance measurement and Voltage/Current monitoring. The Multi-Step function allows on-screen programming of customized measurement sequence with Pass/Fail indication. Each program includes 30 test steps and each test step can be set with selected parameters and test limits. Under Multi-Step operation, a tedious work routine can be done step by step automatically just at a press of a button. With Graph Mode, LCR-8110G, LCR-8105G and LCR-8101G display the component impedance response either over a wide range of test frequency sweep or over a wide range of test voltage sweep in a graph chart. This gives an analysis result of either impedance vs. frequency or impedance vs. applied voltage all at a glance. GPIB and RS-232C interfaces are available as standard for instrument control and test result display on the PC. The rich features of LCR-8000G Series easily make your measurement tasks done at a very competitive price.
TEST FREQUENCY | |
20Hz ~ 10 / 5 / 1 MHz, 5 Digits, ±0.005% | |
INPUT IMPEDANCE | |
100Ω | |
BASIC ACCURACY (*) | |
±0.1% (R, Z, X, G, Y, B, L, C) | |
TEST SPEED | |
AC (> 2kHz) MAX: 75mS FAST: 150mS MEDIUM: 450mS SLOW: 600mS | DC MAX: 30mS FAST: 60mS MEDIUM: 120mS SLOW: 900mS |
TEST SIGNAL LEVELS | |
≦3MHz:10mV~2Vrms, 1mV/Step or 10mV/Step, 2%±5mV >3MHz:10mV~1Vrms, 1mV/Step or 10mV/Step, 2%±5mV | |
SHORT CIRCUIT CURRENT | |
Max. 20mA | |
MEASUREMENT RANGES | |
R, Z, X | 0.1mΩ ~ 100MΩ |
G, Y, B | 10nS ~ 1000S |
L | 0.1nH ~ 100kH |
C | 0.01pF ~ 1F |
D | 0.00001 ~ 9.9999 |
Q | 0.1 ~ 9999.9 |
θ | -180° ~ +180° |
Rdc | 0.1mΩ ~ 100MΩ |
MEASUREMENT PARAMETERS | |
Impedance (Z), Phase Angle (θ), Inductance (L), Capacitance (C), AC Resistance (Rac), Quality Factor (Q), Dissipation Factor (D), Admittance (Y), Conductance (G), Reactance (X), Susceptance (B), DC Resistance (Rdc) | |
SERIES OR PARALLEL EQUIVALENT CIRCUIT | |
C + R, C + D, C + Q, L + R, L + Q, L + D | |
SERIES EQUIVALENT CIRCUIT ONLY | |
X + R, X + D, X + Q | |
PARALLEL EQUIVALENT CIRCUIT ONLY | |
C + G, B + G, B + D, B + Q, B + R, L + G | |
POLAR FORM | |
Z + Phase Angle, Y + Phase Angle | |
AVERAGE | |
1 ~ 256 TIMES | |
LCD DISPLAY | |
320 x 240 DOT-MATRIX | |
INTERFACE | |
RS-232, GPIB | |
POWER SOURCE | |
AC 115V(+10% / -25%), 230V(+15% / -14%) (Selectable), 50/60Hz | |
DIMENSIONS & WEIGHT | |
330(W) x 170(H) x 340(D)mm, Approx. 5kg |
(*):Basic accuracy varies with the speed, frequency, AC signal level and impedance of the device under test.