Search

Categories

  New 
  Used 

Manufacturers

GW Instek GPT-9612

100VA, AC/DC/IR Electrical Safety Tester

More details

1 216,47 €

989,00 € tax excl.

Features

  • AC 100VA AC Withstanding Voltage Tester
  • 100VA AC Test Capacity
  • 240x48 Ice Blue Dot Matrix LCD
  • Auto W-I and I-W Mode
  • True RMS Current Measurement
  • Zero Crossing Turn-on Operation
  • Safety INTERLOCK Function
  • PWM Switching Amplifier to Enhance the Power Efficiency and Reliable Testing   
  • Automatically Switching Input Source for Universal Input Voltage
  • Interface : Remote Terminal, Signal I/O

 

Selection Guide of the GPT-9600 Series
ModelGPT-9601GPT-9602GPT-9612GPT-9603
Output Capacity100VA
AC Withstanding (ACW)5kVac
20mA
5kVac
20mA
5kVac
20mA
5kVac
20mA
DC Withstanding (DCW)------6kVdc
6mA
------6kVdc
6mA
Insulation Resistance (IR)------------50 /100/250/
500/1000Vdc
2000MΩ
50/100/250/
500/1000Vdc
2000MΩ
Ground Bond (GB)------------------------
ARC DetectionYESYESYESYES
Ramp-Up Function0.1s Fixed0.1s Fixed0.1s Fixed0.1s Fixed
Memory Storage for Setting1223
AUTO Mode------------AC-IR or IR-ACAC-IR or IR-AC
DC-IR or IR-DC
Safety InterlockYESYESYESYES
Remote TerminalYESYESYESYES
Signal I/OYESYESYESYES
USB/RS232C Interface------------------------
GPIB Interface------------------------
Display, LCD240 x 48240 x 48240 x 48240 x 48
CE ComplianceYESYESYESYES
AC Withstanding Voltage [for all models]
Output-Voltage Range0.10kV~ 5.00kV
Output-Voltage Resolution10V
Maximum Rated Load100 VA (5kV/20mA)
Maximum Rated Current20mA (0.5kV
Output-Voltage WaveformSine wave
Output-Voltage Frequency50 Hz / 60 Hz selectable
Voltage Regulation± (1.5% + 2 counts) [full load ->no load]
Voltmeter Accuracy± (1.5% of reading + 2 counts)
Current Measurement Range [1]0.01mA~20.0mA
Current Best Resolution0.01mA(0.01mA~9.99mA)
0.1mA(10.0mA~20.0mA)
Current Measurement Accuracy± (2.0% of rdg + 5 counts) when HI SET<1.00mA
± (2.0% of rdg + 3 counts) when HI SET=1.00mA
Window Comparator MethodYes
ARC DetectYes
RAMP (Ramp-Up Time)0.1s Fixed
TIMER (Test Time)OFF, 1s~180s
GNDON
[ 1 ] Output Limitation in AC Withstanding Voltage Testing as below.
 Upper CurrentPauseOutput Time
AC15mA?I?20mAAt least as long as the output timeApprox. 180 seconds
0.01mA?I<15mANot necessaryContinuous output possible
DC Withstanding Voltage [for GPT-9602 / 9603]
Output-Voltage Range0.10kV~ 6.00kV
Output-Voltage Resolution10V
Maximum Rated Load25W (5kV/5mA)
Maximum Rated Current6mA (0.5kV
Voltage Regulation± (1.5% + 2 counts) [full load ->no load]
Voltmeter Accuracy± (1.5% of reading + 2 counts)
Current Measurement Range [1]0.01mA~6.00mA
Current Best Resolution0.01mA(0.01mA~6.00mA)
Current Measurement Accuracy± (2.0% of rdg + 5 counts) when HI SET<1.00mA
± (2.0% of rdg + 3 counts) when HI SET=1.00mA
Window Comparator MethodYes
ARC DetectYes
RAMP (Ramp-Up Time)0.1s Fixed
TIMER (Test Time)OFF, 1s~180s
GNDON
[ 1 ] Output Limitation in DC Withstanding Voltage Testing as below.
 Upper CurrentPauseOutput Time
DC0.01mA?I?6.00mANot necessaryContinuous output possible
Insulation Resistance[for GPT-9612 / 9603]
Output Voltage50V/100V/250V/500V/1000V
Output-Voltage Accuracy± (3.0% of setting + 1 count) with no load
Voltage Regulation± (1.5% + 2 counts) [full load ->no load]
Voltmeter Accuracy± (1.5% of reading + 2 counts)
Resistance Measurement Range1MO~ 2000MO
Test VoltageMeasure RangeAccuracy
50V / 100V / 250V1~50MΩ
51~2000MΩ
± (5% of reading + 2MΩ)
± (10% of reading + 2MΩ)
500V / 1000V1~500MΩ
501~2000MΩ ± (5% of reading + 2MΩ)
± (10% of reading + 2MΩ)
Output Impedance600kΩ
Window Comparator MethodYes
RAMP (Ramp-Up Time)0.1s Fixed
TIMER (Test Time)1s~180s
GNDOFF
General
warning-symbolNoteThe specifications apply when the GPT-9600 series are powered on for at least 30 minutes at 15°C~35°C.
Display240 x 48 Ice Blue dot matrix LCD
AUTO Mode*AC-IR / IR-AC / DC-IR / IR-DC
InterfaceRemote, Signal I/O
Power SourceAC 100-120 V / 220-240 V ±10%
Power Line Frequency50/60Hz
Power ConsumptionMax 400VA
Dimensions322(W) x 148(H) x 385(D) mm
WeightApprox. 9kg max.
* The available Auto mode depends on the selected model
Environment
warning-symbolNoteTo operate over the specific temperature range may cause damage to the circuit. And, do not use in a place where strong magnetic or electric field exists.
Warranty Environment15°C to 35°C, Humidity 70% (no condensation)
Operation Environment0°C to 40°C, Humidity 70% (no condensation)
Storage Environment-10°C to 70°C, Humidity 85% (no condensation)
Installation LocationIndoors at an amplitude of up to 2000m

 