GW Instek GFG-3015
Programmable function generator GFG-3015 15MHz
Search
Manufacturers
Programmable function generator GFG-3015 15MHz
Features
* Frequency Range 0.01Hz~15MHz
* Maximum Frequency Resolution 10mHz
* Dual Displays Indicate Frequency and Amplitude
* Built-in 6 Digits Counter up to 150MHz/High Resolution
* Output Waveforms Sine,Square,Triangle,Ramp,Pulse,Sweep,Gate or Burst
* INT/EXT AM/FM Modulation
* GCV Output for Synchronization
* Output Overload Protection
| |
Output Waveforms | Sine, Square, Triangle, Ramp, Pulse, AM, FM, Sweep, Trigger, Gate or Burst |
Frequency Range | 10mHz~15MHz in 8 frequency range(auto switch) |
Frequency Resolution | 15MHz~150kHz(100Hz) 1.5MHz~15kHz(10Hz) 150kHz~1.5kHz(1Hz) 15kHz~150Hz(0.1Hz) 1.5kHz~15Hz(10mHz) 150Hz~1.5Hz(10mHz) 15Hz~0.15Hz(10mHz) 1.5Hz~0.01Hz(10mHz) |
Frequency Accuracy | 0.02% ±5 count |
Output Impedance | 50Ω±10% |
Amplitude Range | 10.00V ~ 0.01V(into 50Ω) 4 amplitude range |Vac peak| + |Vdc| = 5V Resolution : 10mV (10.00V ~ 0.01V) |
Amplitude Accuracy | <3% ± 5count at 10Hz ~ 1MHz <10% ± 5count at 1MHz ~ 15MHz |
Amplitude Unit | Vpp , Vrms , dBm |
DC Offset | -5V ~ 5V (into 50Ω); resolution 10mV |
DC Offset Accuracy | <3%±3count at minimum amplitude |
Duty Control | 80% ~ 20%, ~ 1MHz; resolution 1% |
Duty Accuracy | <1% ~ 1MHz at 50% duty |
Display | Dual display for frequency 6 digits and amplitude 4 digits |
| |
Distortion | <0.5%(-46dBc), 10Hz~ 100kHz <-30dBc, ~15MHz (1Vpp~10Vpp) |
| |
Linearity Error | <1% of full scale output at 100Hz |
| |
Asymmetry | ±1% of period + 3nS |
Rise or Fail Time | <18 nS |
| |
Impedance | 50Ω±10% |
Level | >1Vp-p open circuit |
| |
Types | AM, FM, Sweep, Trigger(int/ext), Gate or Burst (implement by trigger type) |
Waveform | Sine, Square, Triangle, Ramp or Variable Symmetry Pulse |
Rate Frequency Range | 10mHz~10kHz in 3 frequency range (auto switch) |
Rate Frequency Accuracy | 5% ±1 count |
Rate Frequency Resolution | 10.0kHz~0.1kHz(100Hz) 99Hz~1Hz(1Hz) 0.99Hz ~ 0.01Hz(0.01Hz) |
Symmetry | 90% : 10% : 90%; resolution : 1% |
Symmetry Accuracy | ±1 count(<± 1%) |
Output Level | >1Vpp into 10kΩ load |
Sine Wave Distortion | <2% from 10Hz ~ 10kHz |
| |
Depth | 0 ~ 100% |
Modulation Frequency Rate | 0.01Hz ~ 10kHz(INT) ; DC ~ 1MHz(EXT) |
Carries -3dB Bandwidth | <100Hz to >5MHz |
External Sensitivity | <10Vpp for 100% modulation |
| |
Deviation | 0 ~ ±15% |
Modulation Frequency Rate | 0.01Hz ~ 10kHz(INT) ; DC ~ 50kHz(EXT) |
External Sensitivity | <5Vpp for 15% deviation |
| |
Sweep Range | 15MHz ~ 150kHz; 1.5MHz ~ 15kHz; 150kHz ~ 1.5kHz; 15kHz ~ 150Hz; 1.5kHz ~ 15Hz; 150Hz ~ 1.5Hz; 15Hz ~ 0.15Hz; 1.5Hz ~ 0.01Hz |
Width | >100:1(in the same frequency range) |
Rate | 0.01Hz ~ 10kHz |
Symmetry Control | 90% ~ 10%, resolution:1% |
Mode | Linear /Log sweep |
Sweep Output | 0>-5Vpp into 10kΩ |
| |
Start/Stop Phase Range | 80° ~ 90° |
Rate | 0.1Hz ~ 10kHz |
Frequency Range | 0.1Hz ~ 1MHz(Useful ~ 10MHz) |
Ext Trig Frequency Range | DC ~ 1MHz, TTL compatible input level |
Gate or Burst | Implement by trigger setting |
| |
Range | 100:1 (0 ~ 10V ±1V) |
Input Linearity | <0.5% ~ 1MHz, <5% ~ 10MHz |
Input Impedance | 10 kΩ |
| |
0.2V ~ 2V as per different frequencies in the same frequency range | |
| |
INT/EXT | Switch selector |
Range | 5Hz ~ 150MHz EXT |
Accuracy | Time base accuracy ±1 count |
Time Base | ±20ppm(23°C ±5°C) after 30 minutes warm up |
Resolution | 100nHz for 0.1Hz, 1Hz for 100MHz |
Input Impedance | 1MΩ // 150pF |
Sensitivity | <35mVrms(5 ~ 100MHz), <45mVrms(100MHz~150MHz) |
MEMORY | |
Save/Recall | 10 sets of front panel setting memories |
INTERFACE | |
RS-232C | |
POWER SOURCE | |
AC115V/230V ±15%, 50/60Hz | |
| |
User manual x 1 Power cord x 1 Test Lead GTL-101 x 2 | |
DIMENSIONS & WEIGHT | 290(W) x142(H) x 460(D)mm, Approx. 5kg |