GW Instek PSW 80-13.5
DC Power Supply 360W (80V / 13.5A)
Search
Manufacturers
DC Power Supply 360W (80V / 13.5A)
Model | PSW | 30-36 | 80-13.5 | 160-7.2 | 250-4.5 | 800-1.44 | |
Rated Output Voltage | V | 30 | 80 | 160 | 250 | 800 | |
Rated Output Current | A | 36 | 13.5 | 7.2 | 4.5 | 1.44 | |
Rated Output Power | W | 360 | 360 | 360 | 360 | 360 | |
Power Ratio | -- | 3 | 3 | 3.2 | 3.125 | 3.2 | |
Constant Voltage Mode | PSW | 30-36 | 80-13.5 | 160-7.2 | 250-4.5 | 800-1.44 | |
Line Regulation (*1) | mV | 18 | 43 | 83 | 128 | 403 | |
Load Regulation (*2) | mV | 20 | 45 | 85 | 130 | 405 | |
Ripple and Noise (*3) | |||||||
p-p (*4) | mV | 60 | 60 | 60 | 80 | 150 | |
r.m.s (*5) | mV | 7 | 7 | 12 | 15 | 30 | |
Temperature coefficient | ppm/°C | 100ppm/°C of rated output voltage, after a 30 minute warm-up. | |||||
Remote sense compensation voltage | V/wire | 0.6 | 0.6 | 0.6 | 1 | 1 | |
Rise Time (*6) | |||||||
Rated Load | ms | 50 | 50 | 100 | 100 | 150 | |
No Load | ms | 50 | 50 | 100 | 100 | 150 | |
Fall Time (*7) | |||||||
Rated Load | ms | 50 | 50 | 100 | 150 | 300 | |
No Load | ms | 500 | 500 | 1000 | 1200 | 2000 | |
Transient response time (*8) | ms | 1 | 1 | 2 | 2 | 2 | |
Constant Current Mode | PSW | 30-36 | 80-13.5 | 160-7.2 | 250-4.5 | 800-1.44 | |
Line regulation (*1) | mA | 41 | 18.5 | 12.2 | 9.5 | 6.44 | |
Load regulation (*9) | mA | 41 | 18.5 | 12.2 | 9.5 | 6.44 | |
Ripple and noise | |||||||
r.m.s (*5 ) | mA | 72 | 27 | 15 | 10 | 5 | |
Temperature coefficient | ppm/°C | 200ppm/°C of rated output current, after a 30 minute warm-up. | |||||
Protection Function | PSW | 30-36 | 80-13.5 | 160-7.2 | 250-4.5 | 800-1.44 | |
Over voltage protection (OVP) | |||||||
Setting range | V | 3-33 | 8-88 | 16-176 | 20-275 | 20-880 | |
Setting accuracy | ± (2% of rated output voltage) | ||||||
Over current protection (OCP) | |||||||
Setting range | A | 3.6-39.6 | 1.35-14.85 | 0.72-7.92 | 0.45-4.95 | 0.144-1.584 | |
Setting accuracy | ± (2% of rated output current) | ||||||
Over temperature protection (OTP) | |||||||
Operation | Turn the output off. | ||||||
Low AC input protection (AC-FAIL) | |||||||
Operation | Turn the output off. | ||||||
Power limit (POWER LIMIT) | |||||||
Operation | Over power limit. | ||||||
Value (fixed) | Approx. 105% of rated output power | ||||||
Front Panel | PSW | 30-36 | 80-13.5 | 160-7.2 | 250-4.5 | 800-1.44 | |
Display, 4 digits | |||||||
Voltage accuracy 0.1% + | mV | 20 | 20 | 100 | 200 | 400 | |
Current accuracy 0.1% + | mA | 40 | 20 | 5 | 5 | 2 | |
Programming and Measurement (Interface) | PSW | 30-36 | 80-13.5 | 160-7.2 | 250-4.5 | 800-1.44 | |
Voltage programming accuracy 0.1% + | mV | 10 | 10 | 100 | 200 | 400 | |
Current programming accuracy 0.1% + | mA | 30 | 10 | 5 | 5 | 2 | |
Voltage programming resolution | mV | 1 | 2 | 3 | 5 | 14 | |
Current programming resolution | mA | 1 | 1 | 1 | 1 | 1 | |
Voltage measurement accuracy 0.1% + | mV | 10 | 10 | 100 | 200 | 400 | |
Current measurement accuracy 0.1% + | mA | 30 | 10 | 5 | 5 | 2 | |
Voltage measurement resolution | mV | 1 | 2 | 3 | 5 | 14 | |
Current measurement resolution | mA | 1 | 1 | 1 | 1 | 1 | |